Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4361
Title: High-rate operation of the Micro-Hole and Strip Plate gas detector
Authors: Veloso, J. F. C. A. 
Caldas, C. C. 
Oliveira, C. A. B. 
Azevedo, C. D. R. 
Santos, J. M. F. dos 
Breskin, A. 
Chechik, R. 
Keywords: Count rate
Issue Date: 2007
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 580:1 (2007) 362-365
Abstract: We report the performance of the Micro-Hole-and-Strip-Plate (MHSP) operated in xenon, under high-flux irradiation of a 30-kV X-ray Cu-tube. Stable operation in pulse-counting mode for gas gains above 104 was measured up to counting rates of 5×105 Hz/mm2. A gain variation of less than 5% was observed for gains of 104, while the energy resolution for 8-keV X-rays degrades from ~16% to ~18%. No visible degradation was observed on the MHSP irradiated area.
URI: https://hdl.handle.net/10316/4361
DOI: 10.1016/j.nima.2007.05.179
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
file1cf8cc57e48a40cd87535b6e690522ce.pdf1.04 MBAdobe PDFView/Open
Show full item record

SCOPUSTM   
Citations

8
checked on Oct 28, 2024

WEB OF SCIENCETM
Citations

7
checked on Oct 2, 2024

Page view(s) 50

461
checked on Oct 29, 2024

Download(s) 50

581
checked on Oct 29, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.