Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4456
Title: The X-ray performance of a driftless gas proportional scintillation counter using short shaping-time constants for pulse analysis
Authors: Covita, D. S. 
Simões, P. C. P. S. 
Fernandes, L. M. P. 
Freitas, E. D. C. 
Ferreira, L. F. Requicha 
Veloso, J. F. C. A. 
Santos, J. M. F. dos 
Issue Date: 2004
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 516:1 (2004) 134-142
Abstract: Performance characteristics are evaluated for a xenon driftless gas proportional scintillation counter (GPSC), for which detector pulses are formatted using very short shaping-time constants (~50 ns), and directly analyzed by the MCA without previous pulse time duration analysis and amplitude correction. The present detector and method allow the achievement of detector energy linearity and energy resolution similar to those of conventional GPSCs, reducing background levels and maximizing the detector count-rate capability. However, the obtained peak tails can be somewhat larger than those obtained with conventional GPSCs for X-ray energies above ~30 keV. Energy resolutions of 7.7%, 4.5% and 3.8% can be achieved for 5.9, 22.1 and 59.6 keV X-rays, respectively.
URI: https://hdl.handle.net/10316/4456
DOI: 10.1016/j.nima.2003.07.061
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais

Files in This Item:
File Description SizeFormat
file3e3dcf86b24a4966bf143506954cfbbc.pdf318.85 kBAdobe PDFView/Open
Show full item record

SCOPUSTM   
Citations

3
checked on Nov 9, 2022

WEB OF SCIENCETM
Citations

2
checked on May 2, 2023

Page view(s) 50

473
checked on Oct 30, 2024

Download(s)

285
checked on Oct 30, 2024

Google ScholarTM

Check

Altmetric

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.