Title: | Energy linearity response of CZT detectors to X-rays in the region of the Zn, Cd and Te K-absorption edges: experimental results | Authors: | Freitas, E. D. C. Cardoso, João M. Fernandes, L. M. P. Lopes, J. A. M. Veloso, J. F. C. A. Santos, J. M. F. dos |
Keywords: | Energy linearity; K-edges; X-ray detectors | Issue Date: | 2003 | Citation: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 515:3 (2003) 618-623 | Abstract: | The response of a cadmium zinc telluride (CZT) detector to X-rays with energies around the Zn, Cd and Te K-absorption edges was investigated. The energy non-linearity in the detector response at the K-edges is less than 0.1%, within the experimental errors. No abrupt variation in both the w-value and the detector energy resolution was observed at the K-edges, a behaviour consistent with the absence of intrinsic non-linearity effects in CZT. Within the accuracy of our measurements, we conclude that there are no non-linearity effects in CZT semiconductors at the Zn, Cd and Te K-edges. | URI: | https://hdl.handle.net/10316/4461 | DOI: | 10.1016/j.nima.2003.06.005 | Rights: | openAccess |
Appears in Collections: | FCTUC Física - Artigos em Revistas Internacionais |