Title: | Low-energy calibration of XENON1T with an internal 37 Ar source |
Authors: | Aprile, E. Abe, K. Agostini, F. Ahmed Maouloud, S. Alfonsi, M. Althueser, L. Andrieu, B. Angelino, E. Angevaare, J. R. Antochi, V. C. Antón Martin, D. Arneodo, F. Baudis, L. Baxter, A. L. Bellagamba, L. Biondi, R. Bismark, A. Brown, A. Bruenner, S. Bruno, G. Budnik, R. Bui, T. K. Cai, C. Capelli, C. Cardoso, J. M. R. Cichon, D. Colijn, A. P. Conrad, J. Cuenca-García, J. J. Cussonneau, J. P. D’Andrea, V. Decowski, M. P. DiGangi, P. Di Pede, S. Diglio, S. Eitel, K. Elykov, A. Farrell, S. Ferella, A. D. Ferrari, C. Fischer, H. Fulgione, W. Gaemers, P. Gaior, R. Gallo Rosso, A. Galloway, M. Gao, F. Glade-Beucke, R. Grandi, L. Grigat, J. Guida, M. Hammann, R. Higuera, A. Hils, C. Hoetzsch, L. Howlett, J. Iacovacci, M. Itow, Y. Jakob, J. Joerg, F. Joy, A. Kato, N. Kara, M. Kavrigin, P. Kazama, S. Kobayashi, M. Koltman, G. Kopec, A. Kuger, F. Landsman, H. Lang, R. F. Levinson, L.. Li, I. Li, S. Liang, S. Lindemann, S. Lindner, M. Liu, K. Loizeau, J. Lombardi, F. Long, J. Lopes, J. A. M. Ma, Y. Macolino, C. Mahlstedt, J. Mancuso, A. Manenti, L. Marignetti, F. Marrodán Undagoitia, T. Martens, K. Masbou, J. Masson, D. Masson, E. Mastroianni, S. Messina, M. Miuchi, K. Mizukoshi, K. Molinario, A. Moriyama, S. Morå, K .D. Mosbacher, Y. Murra, M. Müller, J. Ni, K. Oberlack, U. Paetsch, B. Palacio, J. Peres, R. Peters, C. Pienaar, J. Pierre, M. Pizzella, V. Plante, G. Qi, J. Qin, J. Ramírez García, D. Reichard, S. Rocchetti, A. Rupp, N. Sanchez, L. Sanchez-Lucas, P. Santos, Joaquim Marques Ferreira dos Sarnoff, I. Sartorelli, G. Schreiner, J. Schulte, D. Schulte, P. Schulze Eißing, H. Schumann, M. Lavina, L. Scotto Selvi, M. Semeria, F. Shagin, P. Shi, S. Shockley, E. Silva, M. Simgen, H. Takeda, A. Tan, P.-L. Terliuk, A. Thers, D. Toschi, F. Trinchero, G. Tunnell, C. Tönnies, F. Valerius, K. Volta, G. Weinheimer, C. Weiss, M. Wenz, D. Wittweg, C. Wolf, T. Xu, D. Xu, Z. R. Yamashita, M. Yang, L. Ye, J. Yuan, L. Zavattini, G. Zerbo, S. Zhong, M. Zhu, T. Geppert, C. Riemer, J. |
Issue Date: | 2023 |
Publisher: | Institute for Ionics |
Serial title, monograph or event: | European Physical Journal C |
Volume: | 83 |
Issue: | 6 |
Abstract: | A low-energy electronic recoil calibration of
XENON1T, a dual-phase xenon time projection chamber,
with an internal 37Ar source was performed. This calibration
source features a 35-day half-life and provides two monoenergetic
lines at 2.82 keV and 0.27 keV. The photon yield
and electron yield at 2.82 keVare measured to be (32.3±0.3)
photons/keV and (40.6 ± 0.5) electrons/keV, respectively, in
agreement with other measurements and with NEST predictions.
The electron yield at 0.27 keVis alsomeasured and it is
(68.0+6.3
−3.7) electrons/keV. The 37Ar calibration confirms that
the detector is well-understood in the energy region close to
the detection threshold, with the 2.82 keV line reconstructed
at (2.83 ± 0.02) keV, which further validates the model used
to interpret the low-energy electronic recoil excess previously
reported by XENON1T. The ability to efficiently remove
argon with cryogenic distillation after the calibration proves
that 37Ar can be considered as a regular calibration source
for multi-tonne xenon detectors. |